Page 230 - BOLA Catalogue
P. 230

   BOLA Temperature Probes and Total Immersion Probes
Build-up and function of BOLA Temperature Probes
Two different sensors are used for BOLA Temperature Probes.
BOLA Temperature Probes PT 100 and PT 1000 are resistance thermometers which measure temperature correlated to platinum‘s changing electrical resistance under temperature influence with a deposited table of values.
BOLA Temperature Probes Type K are thermocouples which measure temperature with the help of changing voltage inside the sensor under temperature influence.
The sensors of all BOLA Temperature Probes, PT 100, PT 1000 and Type K are located at the end of a PTFE-encapsulated stainless steel tube (material code: 1.4571).
The stainless steel tube provides certain rigidity, but can be bent to
the requested form by hand, so that the sensor can be oriented to the optimum measuring point.
For BOLA Temperatures Probes PT 100 and PT 1000, connection to the measuring device is made either with a socket or plug type Lemo® size 1 or by connecting directly the strands of the cable to the device.
For BOLA Temperature Probes Type K, connection to the measuring device is either made by using the SMP-connector or by connecting directly the strands to the device.
The cable itself is also encapsulated with PFA and connected tightly to the temperature probe
Advantages of BOLA Temperature Probes Reduction of response time
Temperature Probes PT 100 and PT 1000 have tapered tips which reduce the response time considerably.
Chemical resistance and metal-free
The PTFE encapsulation provides an almost universal chemical resistance. All parts which are exposed to the medium do not contain any metals.
Safe to handle
Due to the collar ring at its end, the probe cannot fall into the medium.
High accuracy of measurement
The four-wire system of the PT 100 probes eliminates nearly completely the influence of the resistivity and the transition resistance (failure: approx. 0,002-0,004 % / Ohm). Normally, a calibration is not necessary. Long connecting cables can be used with the four-wire system.
PT 1000 probes use the two-wire system. At 0 °C, PT 100 probes have a resistance of 100 Ohm whereas PT 1000 probes have a resistance of 1000 Ohm. Due to the high inherent resistance of PT 1000 probes, the measuring result is insignificantly influenced by the resistivity of the measuring line. Thus the measured temperature deviates minimally from the actual temperature even when long connecting cables are used.
Performance Data of BOLA Temperature Probes 4-wire-system - PT 100
Pin configuration of the LEMO® socket/plug
TEMPERATURE MEASUREMENT
 By the way: BOLA offers custom manufacture. From 1 piece. Hotline +49 (0) 93 46-92 86-0
229
Temperature probe / PT 100 Temperature range: Specification:
Type:
Class:
Tolerance:
Typical aberrations:
-50°C to +250°C
DIN EN 60751
Platinum temperature sensor A
0,15 + (0,002 x (t))
at 0°C: +/-0,15°C
at 100°C: +/- 0,35°C